dc.contributor.author | Canty, W. J. | en_US |
dc.date.accessioned | 2009-05-08T05:21:28Z | |
dc.date.available | 2009-05-08T05:21:28Z | |
dc.date.issued | 1952-12-02 | en_US |
dc.identifier | MC665_r05_M-1736.pdf | en_US |
dc.identifier.uri | http://hdl.handle.net/1721.3/39338 | |
dc.description.abstract | A trip to General Ceramics and Steatite Corporation is reported. Calibration of the Tektronix 513-D Oscilloscope used in conjunction with ferrite-core pules-test equipment showed the internal voltage calibrator to be accurate in the region of 0.2 v while the sweep speed calibration was off as much as 20 per cent on some ranges. Results of tests made on the ferrite-core tester at MIT and repeated on the at General Ceramics showed excellent agreement. A new type test jig in which cores to be tested are mounted on a piece of brass tubing gives results the same as the old test jig. Tests on a sample of new MF-1326B, and F-291, cores showed much better uniformity than previous cores of this type. | en_US |